Home / Recalls / Medical Devices
Ameditech Inc recalls QuickTox Drug Screen DipCard(AMT Item/ AMT Description): QT70/ "QuickTox¿ Drug Screen Dipcard 11 Drug Detection: COC3
Recalled October 7, 2020 · FDA recall Z-3043-2020 · Ameditech
Hazard
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling. · FDA Class II
Units
170,425 units
Sold at
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Description
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Product
QuickTox Drug Screen DipCard(AMT Item/ AMT Description): QT70/ "QuickTox¿ Drug Screen Dipcard 11 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+OXY100+MDMA500", QT88/ "QuickTox¿ Drug Screen Dipcard 12 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+OXY100+ MDMA500+BUP10", QT11 / "QuickTox¿ Drug Screen Dipcard 5 Drug Detection: COC300+OPI2000+THC50+AMP1000+PCP25", QT52/ "QuickTox¿ Drug Screen Dipcard 9 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+MTD300+OXY100", QT44/ "QuickTox¿ Drug Screen Dipcard 8 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+OXY100", QT23/ "QuickTox¿ Drug Screen Dipcard 6 Drug Detection: COC300+OPI2000+MET500+THC50+AMP1000+PCP25", QT80F/ "QuickTox¿ Drug Screen Dipcard 12 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+PCP25+BZO300+BAR300+MTD300+TCA1000+OXY100+MDMA500 with Positive Control Line", QT52A/ "QuickTox¿ Drug Screen Dipcard 9 Drug Detection: COC300+OPI300+MET500+THC50+AMP1000+BZO300+BAR300+MTD300+OXY100 4 Adulteration Test: Cr, Ni, pH, Ox ", QT20/ "QuickTox" Drug Screen Dipcard 5 Drug Detection: COC300+OPI2000+MET500+THC50+AMP1000"
Manufacturer(s)
Ameditech Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →