Amendia, Inc recalls Mac Pin Non-Cannulated Screw, 6.5 x 60mm
Recalled January 17, 2018 · FDA recall Z-0325-2018 · Amendia
Hazard
Incorrect labeling for the lot of MAC Pins. Although the pins are laser as 460 mm in length, the pins are 60 mm in length. · FDA Class II
Units
41 screws
Sold at
Distribution to Georgia, USA.
Description
Incorrect labeling for the lot of MAC Pins. Although the pins are laser as 460 mm in length, the pins are 60 mm in length.
Product
Mac Pin Non-Cannulated Screw, 6.5 x 60mm
Manufacturer(s)
Amendia, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →