RecallWatchUSA

Home / Recalls / Medical Devices

American Contract Systems Inc. recalls Medical convenience kits Item Number/Description RCGY52D GYN LAPAROSCOPY SPH 0246965 UIGN82M GYN (PS 907645) UIHY

Recalled October 22, 2025 · FDA recall Z-0179-2026 · American Contract Systems

Hazard
Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed. · FDA Class II
Units
32,433 total
Sold at
US Nationwide distribution in the states of SD, IA, MN, WA, IL.

Description

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Product
Medical convenience kits Item Number/Description RCGY52D GYN LAPAROSCOPY SPH 0246965 UIGN82M GYN (PS 907645) UIHY16AE ASC HYSTEROSCOPY PACK (PS 020634)
Manufacturer(s)
American Contract Systems Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov