Home / Recalls / Medical Devices
Angiodynamics, Inc. recalls Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601-US, Model # (UPN) H787900601US0, STERILE --- This is only for
Recalled February 14, 2018 · FDA recall Z-0530-2018 · Angiodynamics
Hazard
The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator. · FDA Class II
Units
14,353 units in total
Sold at
Worldwide distribution - US Nationwide
Description
The firm is recalling product that has the potential to not function during use. The product affected by this recall will not deliver the desired microwave energy due to coolant ingress in an electrical connection; this condition results in an error code High Reflective Power displayed on the microwave generator.
Product
Accu2i pMTA Applicator, 29 cm LONG, REF/Catalog # 900-601-US, Model # (UPN) H787900601US0, STERILE --- This is only for use with a Sulis V Generator, REF 806-10. Product Usage: The Acculis Accu2i pMTA Microwave Tissue Ablation Applicator is indicated for the intraoperative coagulation of soft tissue.
Manufacturer(s)
Angiodynamics, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →