RecallWatchUSA

Home / Recalls / Medical Devices

Angiodynamics, Inc. recalls AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012

Recalled September 11, 2019 · FDA recall Z-2444-2019 · Angiodynamics

Hazard
The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used. · FDA Class II
Units
109
Sold at
The products were distributed to the following US states: DC, IL, IN, MI, NJ, TN, TX, and WI.

Description

The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.

Product
AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012
Manufacturer(s)
Angiodynamics, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov