Home / Recalls / Medical Devices
Angiodynamics Inc. (Navilyst Medical Inc.) recalls Bio-Stable 5F SL-55CM IR-145 Kit Valved PG, UPN H965458180, Catalog No. 45-818 The Xcela with PASV, BioFlo PICC with
Recalled December 20, 2017 · FDA recall Z-0244-2018 · Angiodynamics Inc. (Navilyst Medical Inc.)
Description
Navilyst Medical, Inc. (NMI) the manufacturer of the Xcela PICC with PASV, BioFlo PICC with PASV and the BioFlo PICC, conducted this recall to the end user level based on information received from Greatbatch Medical, the manufacturer of the ViaPeel PTFE Peelable Introducer. Greatbatch Medical determined that the products listed in their 11/11/2016 Recall Notification have the potential for the handles to detach from the sheath during use. NMI has confirmed that the affected sheaths, Greatbatch Model Number 10890-006, had been included in packaged Xcela and BioFlo PICC Kits.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →