Applied Medical Resources Corp recalls Applied Medical's Inzii¿ 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is class

Recalled November 20, 2013 · FDA recall Z-0183-2014 · Applied Medical Resources

Hazard
Applied Medical is conducting a voluntary recall of the Inzii¿ 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the sterile barrier. The likelihood of this situation to occur and result in patient harm is highly unlikely; however, out of an abundance of caution for patient safety and a commitmen · FDA Class II
Units
29,506 units
Sold at
Worldwide Distribution: US (nationwide) and countries of: Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Japan, Lebanon, Malaysia, Mexico, New Zealand, Netherlands, Norway, Peru, Poland, Portugal, Saudi Arabia, Singapore, Spain, South Africa, Sweden, Switzerland, United Arab Emirates, and United Kingdom.

Description

Applied Medical is conducting a voluntary recall of the Inzii¿ 12/15mm retrieval system. During shipment, the retrieval system packaging has the potential to become punctured with small holes, which could compromise the sterile barrier. The likelihood of this situation to occur and result in patient harm is highly unlikely; however, out of an abundance of caution for patient safety and a commitmen

Product
Applied Medical's Inzii¿ 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is classified as a Class II device. It was cleared to market via 510(k) filing K060051. The retrieval bag is a disposable specimen retrieval bag intended for use as receptacle for the collection of tissue, organs, and calculi during laparoscopic procedures. Applieds CD004 has a shelf life of three years. It is provided sterile and packaged in a heat sealed Tyvek pouch.
Manufacturer(s)
Applied Medical Resources Corp

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Applied Medical Resources Corp recalls Applied Medical's Inzii¿ 12/15mm Retrieval System. Applieds Inzii Tissue Retrieval Bag, Model Number CD004, is class | RecallWatchUSA