RecallWatchUSA

Home / Recalls / Medical Devices

Applied Medical Resources Corp recalls DR. THAMES - LAP CHOLE, REF: K0880, UDI: (01)30607915139620

Recalled June 3, 2020 · FDA recall Z-2160-2020 · Applied Medical Resources

Hazard
There may be a slightly protruding rivet on the distal end of the laparoscopic grasper which has the potential for tissue to catch on it during use and lead to tissue damage. · FDA Class II
Units
21 units
Sold at
US: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY OUS: Argentina, Australia, Austria, Belgium, Canada, Chile, Costa Rica, Denmark, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Latvia, Liechtenstein, Luxembourg, Netherlands, New Zealand, Peru, Portugal, Serbia, Singapore, South Africa, Spain, Thailand, United Kingdom

Description

There may be a slightly protruding rivet on the distal end of the laparoscopic grasper which has the potential for tissue to catch on it during use and lead to tissue damage.

Product
DR. THAMES - LAP CHOLE, REF: K0880, UDI: (01)30607915139620
Manufacturer(s)
Applied Medical Resources Corp

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov