RecallWatchUSA

Home / Recalls / Medical Devices

Applied Medical Resources Corp recalls Laparoscope, general; plastic surgery Used for general, abdominal, gynecological and thoracic minimally invasive surg

Recalled January 16, 2013 · FDA recall Z-0644-2013 · Applied Medical Resources

Hazard
Applied Medical is conducting a voluntary recall on a specific lot number of its 12x150mm Kii shielded bladed access systems. This is due to a potential for the blade to remain exposed after shield deployment. Further use of any affected product should cease immediately. The affected model number is CTB71, and the affected lot is 1179107. · FDA Class II
Units
72 units
Sold at
Worldwide Distribution - USA (nationwide) and France.

Description

Applied Medical is conducting a voluntary recall on a specific lot number of its 12x150mm Kii shielded bladed access systems. This is due to a potential for the blade to remain exposed after shield deployment. Further use of any affected product should cease immediately. The affected model number is CTB71, and the affected lot is 1179107.

Product
Laparoscope, general; plastic surgery Used for general, abdominal, gynecological and thoracic minimally invasive surgical procedures to establish a path of entry or to gain access through tissue planes and/or potential spaces for endoscopic instruments.
Manufacturer(s)
Applied Medical Resources Corp

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Applied Medical Resources Corp recalls Laparoscope, general; plastic surgery Used for general, abdominal, gynecological and thoracic minimally invasive surg | RecallWatchUSA