Argon Medical Devices, Inc recalls Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19N, 1/19 GA Flexcore Biopsy Needle, 1.7F Intr

Recalled February 17, 2021 · FDA recall Z-1074-2021 · Argon Medical Devices

Hazard
Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function. · FDA Class II
Units
187 units
Sold at
US: AL AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA,, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX,, UT, VA, WA, WI, WV, WY OUS: Switzerland, Saudi Arabia, Hong Kong, Thailand, Portugal, Taiwan, Austria, Brazil, Netherlands, Italy, Germany, Colombia, Singapore, Canada, View Nam, Slovakia, Spain

Description

Due to a manufacturing error their is a potential that the sheath within the biopsy kit can separate during use if the tip is flexed or experiences bending stresses. This could result in need for surgical intervention to preclude or reverse permanent damage to a body structure or permanent impairment of a body function.

Product
Argon Medical Devices TLAB, Transjugular Liver Biopsy System, REF: TL - 19N, 1/19 GA Flexcore Biopsy Needle, 1.7F Introducer Sheath W/Curved Metal Stiffener, 4/Tissue Removal Swabs, Sterile EO,
Manufacturer(s)
Argon Medical Devices, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov