RecallWatchUSA

Home / Recalls / Medical Devices

Argon Medical Devices, Inc recalls Arterial Line Kit 20ga X 3", Model 498106, Rx Only, Sterile EO, UDI: 00886333215577 - Product Usage: Arterial Catheters

Recalled May 20, 2020 · FDA recall Z-1937-2020 · Argon Medical Devices

Hazard
Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead. · FDA Class II
Units
20
Sold at
US Nationwide distribution including the states of CA, IL, ND, SC, TX, VA. OUS: None

Description

Arterial Line Kits include thicker curved reverse-cutting needle with a 0-silk suture, instead of the usual straight-cutting needle with a 3-0 silk suture. The thicker suture may be more difficult to tie. The curved needle requires the use of a needle driver, there is a risk of needle stick to the surgeon tries to do it with fingers instead.

Product
Arterial Line Kit 20ga X 3", Model 498106, Rx Only, Sterile EO, UDI: 00886333215577 - Product Usage: Arterial Catheters are indicated for use when in instances where continuous invasive blood pressure readings and/or frequent blood gas measurements are required. Case Label: Arterial Line Kit/1 20ga x 3", Qty: 10/CS, CE 2797
Manufacturer(s)
Argon Medical Devices, Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Argon Medical Devices, Inc recalls Arterial Line Kit 20ga X 3", Model 498106, Rx Only, Sterile EO, UDI: 00886333215577 - Product Usage: Arterial Catheters | RecallWatchUSA