Home / Recalls / Medical Devices
Arjo, Inc. dba ArjoHuntleigh recalls Medium 2-Hooks Spreader Bar for use with Loop Slings, sizes small, medium and large on the Maxi Sky 2 Ceiling Lift; Mode
Recalled October 22, 2014 · FDA recall Z-0079-2015 · Arjo, Inc.
Hazard
The upper portion of the hooks on the spreader bar are too flexible for holding the loops of the sling. The loop of the sling could potentially come out of the hooks if positioned in the upper portion of the hooks. · FDA Class II
Units
363 spreader bars
Sold at
USA Distribution - AR, GA, MI, PA, and the District of Columbia.
Description
The upper portion of the hooks on the spreader bar are too flexible for holding the loops of the sling. The loop of the sling could potentially come out of the hooks if positioned in the upper portion of the hooks.
Product
Medium 2-Hooks Spreader Bar for use with Loop Slings, sizes small, medium and large on the Maxi Sky 2 Ceiling Lift; Model 700-19425. Intended to be used for general transfers in hospitals.
Manufacturer(s)
Arjo, Inc. dba ArjoHuntleigh
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →