RecallWatchUSA

Home / Recalls / Medical Devices

Arkray Factory USA, Inc. recalls arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Ure

Recalled January 27, 2016 · FDA recall Z-0631-2016 · Arkray Factory USA

Hazard
One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL. · FDA Class I
Units
11 boxes (25 foil packaged test strips per box)
Sold at
Distributed in the states of FL, IL, KY, MI, NC, NY, OH, and TN.

Description

One lot each of SPOTCHEM II Basic PANEL-1 Reagent Test Strip and SPOTCHEM II Glucose Reagent Test Strip are being recalled due to a lowered reaction rate in the reagent pad for glucose. The affected lots will display a result of approximately 300 to 400 mg/dL for any glucose value that exceeds the upper limit of 416 mg/dL.

Product
arkray SPOTCHEM II Basic Panel - 1 Reagent test strips, Product # 11726, MEDTEST DX Catalog # 77484. Calcium, Blood Urea Nitrogen, Glucose, Albumin, Creatinine. For the in vitro quantitative determination of calcium, blood urea nitrogen (BUN), glucose, albumin, and creatinine.
Manufacturer(s)
Arkray Factory USA, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov