RecallWatchUSA

Home / Recalls / Medical Devices

Arobella Medical, LLC recalls Arobella Medical 6mm Qurette used with the Arobella Medical AR1000 Wound Therapy System. Major components of the AR1

Recalled December 26, 2012 · FDA recall Z-0534-2013 · Arobella Medical

Hazard
This recall has been initiated because the Arobella Medical 6mm Qurette Q07 Lot Numbers 0001 and 0002 can fracture during use and may release a fragment or fragments. The 6 mm Qurette is a component of the AR1000 Quostic Wound Therapy System. · FDA Class II
Units
99
Sold at
Worldwide distribution - US (nationwide) and the states of AL, CA, ID, IN, IA, MI, MN, MS, NJ, NM, OH, PA. SC, TX, UT, VA, WI. including the country of SAUDI ARABIA.

Description

This recall has been initiated because the Arobella Medical 6mm Qurette Q07 Lot Numbers 0001 and 0002 can fracture during use and may release a fragment or fragments. The 6 mm Qurette is a component of the AR1000 Quostic Wound Therapy System.

Product
Arobella Medical 6mm Qurette used with the Arobella Medical AR1000 Wound Therapy System. Major components of the AR1000 Wound Therapy System include an ultrasonic generator, converter hand piece, done-shaped applicator tip (qurette), and a saline supply reservoir. Product Usage: The Arobella Medical AR1000 Wound Therapy System is interned for the selective dissection and fragmentation of tissue, wound debridement ( of acute and chronic wounds, burns, and diseased or necrotic tissue), and cleansing irrigation of the wound site for the removal of debris, exudates, fragments, and other matter.
Manufacturer(s)
Arobella Medical, LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Arobella Medical, LLC recalls Arobella Medical 6mm Qurette used with the Arobella Medical AR1000 Wound Therapy System. Major components of the AR1 | RecallWatchUSA