RecallWatchUSA

Home / Recalls / Medical Devices

Arrow International Inc recalls AF 550 x 5ML/H 6" DUAL CATHS, 4.5", Product code MC0050XLYK5 CP, Arrow AutoFuser Disposable Pain Control Pump

Recalled March 6, 2019 · FDA recall Z-0946-2019 · Arrow International

Hazard
Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration. · FDA Class II
Units
8
Sold at
Nationwide distribution.

Description

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

Product
AF 550 x 5ML/H 6" DUAL CATHS, 4.5", Product code MC0050XLYK5 CP, Arrow AutoFuser Disposable Pain Control Pump
Manufacturer(s)
Arrow International Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Arrow International Inc recalls AF 550 x 5ML/H 6" DUAL CATHS, 4.5", Product code MC0050XLYK5 CP, Arrow AutoFuser Disposable Pain Control Pump | RecallWatchUSA