Home / Recalls / Medical Devices
Arrow International Inc recalls Arrow AutoCAT2 Intra-Aortic Balloon Pump AEROAUTOCAT 2 WAVE, Product Code IAP-0535
Recalled July 8, 2020 · FDA recall Z-2415-2020 · Arrow International
Hazard
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP. · FDA Class I
Units
2123 total
Sold at
Worldwide distribution.
Description
A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.
Product
Arrow AutoCAT2 Intra-Aortic Balloon Pump AEROAUTOCAT 2 WAVE, Product Code IAP-0535
Manufacturer(s)
Arrow International Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →