Home / Recalls / Medical Devices
ARROW INTERNATIONAL Inc. recalls Arrow Ultra 8 IAB Intra-Aortic Balloon Catheter Kit, a) REF IAB-05830-U; b) REF IAB-05840-U
Recalled June 12, 2024 · FDA recall Z-1880-2024 · ARROW INTERNATIONAL
Hazard
Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway. · FDA Class I
Units
3138 units
Sold at
Worldwide distribution. US nationwide including Puerto Rico, AU, BA, BE, BG, BN, CA, CL, CN, CO, CY, CZ, DE, EC, ES, FR, GB, GP, GR, HU, HK, IE, IL, IN, IQ, IT, JO, JP, KW, KZ, LB, LU, MA, MN, MK, MX, MY, NL, OM, PA, PF, PH, PL, PT, QA, RE, RO, RS, SA, SE, SG, SI, SK, TH, TR, TW, UZ, ZA, and UAE.
Description
Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.
Product
Arrow Ultra 8 IAB Intra-Aortic Balloon Catheter Kit, a) REF IAB-05830-U; b) REF IAB-05840-U
Manufacturer(s)
ARROW INTERNATIONAL Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →