RecallWatchUSA

Home / Recalls / Medical Devices

ArthroCare Medical Corporation recalls Coblator II System, 120V. Product Usage: The Coblator II (Model RF8000E) is a bipolar, radiofrequency electrosurgic

Recalled April 1, 2015 · FDA recall Z-1323-2015 · ArthroCare Medical

Hazard
The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E). · FDA Class II
Units
15 units
Sold at
US Nationwide Distribution in the states of TN, IA, AZ, NE, SD, NY, and OK.

Description

The recalling firm was notified by the contract manufacturer that they had reversed the L2, L3, L4, or L5 components on the P/N09010 RoHS boards of the Coblator II Controller (Model RF8000E).

Product
Coblator II System, 120V. Product Usage: The Coblator II (Model RF8000E) is a bipolar, radiofrequency electrosurgical system designed for use in otorhinolaryngology (ENT) surgery. The system consists of the following components: an electrosurgical radiofrequency controller, a reusable, non-sterile Power Cord; a reusable, non-sterile Foot Control; a reusable, non-sterile Flow control Valve Unit; a reusable, non-sterile flow control cable; a single use, disposable, sterile Wand. The Coblator II is indicated for ablation, resection, and coagulation of soft tissue and hemostasis of blood vessels in otorhinolaryngology (ENT) surgery.
Manufacturer(s)
ArthroCare Medical Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

ArthroCare Medical Corporation recalls Coblator II System, 120V. Product Usage: The Coblator II (Model RF8000E) is a bipolar, radiofrequency electrosurgic | RecallWatchUSA