RecallWatchUSA

Home / Recalls / Medical Devices

Asensus Surgical Inc recalls Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Sur

Recalled November 8, 2023 · FDA recall Z-0170-2024 · Asensus Surgical

Hazard
Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed. · FDA Class I
Units
21 units
Sold at
US: NJ, LA, FL, MN, IL, and Germany, Japan, Lithuania

Description

Asensus Surgical has identified a malfunction that has occurred and may recur on the Senhance Surgical System. This issue presented itself as uncontrolled arm motion of the Laparoscope Instrument Actuator (LIA) where the LIA rotated continuously in one direction after the Surgeon removed the engagement of teleoperation on the Senhance System. There was no other uncontrolled arm motion observed.

Product
Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Surgical Unit (Catalog Number X9007696), and Cockpit, ISU Configuration (Catalog Number X9007707), Software (SW) version 2.7.4; robotic surgical system
Manufacturer(s)
Asensus Surgical Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Asensus Surgical Inc recalls Senhance Surgical System, which includes: Manipulator Arm, ISU Configuration ( Catalog Number X9007708), Intelligent Sur | RecallWatchUSA