Home / Recalls / Medical Devices
Auris Health, Inc recalls MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-00000
Recalled February 25, 2026 · FDA recall Z-1383-2026 · Auris Health
Hazard
Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube. · FDA Class II
Units
173
Sold at
Worldwide - US Nationwide distribution in the states of MN, FL, NJ, NY, VA, WI, AZ, AL, MS, LA, MI, MA, MT, PA, WV, GA, DE, TX, CA, OH, IN, SC, CT, ND, MO, DC, IL, NE, ID, WA, NC, TN, OR, CO, SD, NM, VT, MD and the countries of Hong Kong, China, Canada.
Description
Device for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.
Product
MONARCH Platform, Bronchoscopy, Catalog Numbers: MON-000005-01 , MON-000005-01R , MON-000006 , MON-000006-RFB, MON-000008 with software
Manufacturer(s)
Auris Health, Inc
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →