RecallWatchUSA

Home / Recalls / Medical Devices

AURIS HEALTH INC recalls Monarch Platform (Endoscopy), REF: MON-000005-01, MON-000005-01R. Monarch Platform (Bronchoscopy), REF: MON-000006, M

Recalled October 16, 2024 · FDA recall Z-0051-2025 · AURIS HEALTH

Hazard
Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock. · FDA Class II
Units
176
Sold at
Worldwide - US Nationwide distribution including in the states of MN, FL, NC, AK, NJ, NY, VA, WI, AZ, MS, LA, MI, MA, MT, WV, GA, DE, TX, CA, OH, IN, SC, PA, CT, ND, MO, DC, IL, NE, ID, WA, TN, OR, AL, SD, CO, KY, NM, VT, MD and the countries of Canada.

Description

Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.

Product
Monarch Platform (Endoscopy), REF: MON-000005-01, MON-000005-01R. Monarch Platform (Bronchoscopy), REF: MON-000006, MON-000006-RFB.
Manufacturer(s)
AURIS HEALTH INC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

AURIS HEALTH INC recalls Monarch Platform (Endoscopy), REF: MON-000005-01, MON-000005-01R. Monarch Platform (Bronchoscopy), REF: MON-000006, M | RecallWatchUSA