AVID Medical, Inc. recalls Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK

Recalled August 12, 2026 · FDA recall Z-2796-2026 · AVID Medical

Hazard
Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels. · FDA Class I
Units
5168 kits
Sold at
US Nationwide distribution in the states of NC and TX.

Description

Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.

Product
Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05
Manufacturer(s)
AVID Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov