B. Braun Medical, Inc. recalls Diacap(R) Ultra Dialysis Fluid Filter, Catalog Number 7107367. Intended to filter bacteria and endotoxins from dialys
Recalled May 8, 2013 · FDA recall Z-1183-2013 · B. Braun Medical
Hazard
There is the potential of the residual moisture of the Diacap Ultra membrane being reduced over time, which results in the reduction of the membrane permeability (ultrafiltration coefficient). · FDA Class II
Units
1580
Sold at
Distribution and manufacturing dates for the expansion lots was reported as 5/23-6/14/2013 and 1/22-24/2013 respectively. Nationwide Distribution including the states of AZ, CA, DE, FL, GA, HI, IA, ID, IL, KS, KY, LA, MN, MO, NJ, NM, NY, OH, OR, PA, SC, SD, TN, TX, UT, WA and WV.
Description
There is the potential of the residual moisture of the Diacap Ultra membrane being reduced over time, which results in the reduction of the membrane permeability (ultrafiltration coefficient).
Product
Diacap(R) Ultra Dialysis Fluid Filter, Catalog Number 7107367. Intended to filter bacteria and endotoxins from dialysis fluid.
Manufacturer(s)
B. Braun Medical, Inc.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →