B. Braun Medical, Inc. recalls Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per carton. A6 Access Device is a valve intended

Recalled October 26, 2022 · FDA recall Z-0091-2023 · B. Braun Medical

Hazard
The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia. · FDA Class II
Units
354,200 US; 1,800 OUS
Sold at
Distribution US nationwide, Canada and Saudi Arabia.

Description

The extension set has a high probability of failure which could lead to possible leakage or detachment at the device's bonded joint during use. The issue may result in delay of therapy, increased risk of bloodstream infections (BSI), and may expose both patients and healthcare workers to hazardous medications. Depending on the nature of the leak and type of infusion, there is also the potential for life-threatening situations to occur including air embolization or hypovolemia.

Product
Extension Set, EXT SET W/CARESITE, 6IN., Product Code 470169, 50 units per carton. A6 Access Device is a valve intended for the aspiration, injection or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector.
Manufacturer(s)
B. Braun Medical, Inc.