RecallWatchUSA

Home / Recalls / Medical Devices

B Braun Medical Inc recalls Infusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U

Recalled September 18, 2024 · FDA recall Z-2972-2024 · B Braun Medical

Hazard
Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected. · FDA Class I
Units
14
Sold at
US distribution to AL, CA, GA, IL, MI, and TX.

Description

Extension of previous recall RES 92978. Pumps have the potential for upstream occlusion sensor to cease proper function due to buildup of electrostatic charge during operation triggering false downstream and upstream occlusion alarms. Additional devices have been affected.

Product
Infusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U
Manufacturer(s)
B Braun Medical Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

B Braun Medical Inc recalls Infusomat Space Infusion System/ Large Volume Pump, Product Code 8713050U | RecallWatchUSA