Home / Recalls / Medical Devices
B. Braun Medical, Inc. recalls Outlook Safety Infusion System Burette Set. For intravenous fluid administration.
Recalled September 18, 2013 · FDA recall Z-2166-2013 · B. Braun Medical
Hazard
Reports were received of an incorrectly assembled Outlook Safety Infusion System Burette Set (Product Code 375038), and the investigation determined that one of the ULTRASITE injection sites and corresponding low pressure backcheck valve was assembled upside down. · FDA Class II
Units
900
Sold at
Distributed in the states of FL, NY, NJ, and TX.
Description
Reports were received of an incorrectly assembled Outlook Safety Infusion System Burette Set (Product Code 375038), and the investigation determined that one of the ULTRASITE injection sites and corresponding low pressure backcheck valve was assembled upside down.
Product
Outlook Safety Infusion System Burette Set. For intravenous fluid administration.
Manufacturer(s)
B. Braun Medical, Inc.