RecallWatchUSA

Home / Recalls / Medical Devices

B. Braun Medical, Inc. recalls STOPCOCK EXT. SET W/ 3-WAY STOPCOCK WITH INJ. SITE AND SPINLOCK CONNECTOR, 22 inches and 37 inches

Recalled July 16, 2014 · FDA recall Z-2022-2014 · B. Braun Medical

Hazard
Complaints were received that reported the stopcock of some Stopcock Extension Sets are assembled in the reverse position, resulting in the stopcock being closed when set in the open position and vise versa. · FDA Class II
Units
10,835
Sold at
USA (nationwide) including the states of PA, IL, NC, FL, IL, CA, CO, NE, LA, NC, LA, VA, TN, NE, TX, and KS.

Description

Complaints were received that reported the stopcock of some Stopcock Extension Sets are assembled in the reverse position, resulting in the stopcock being closed when set in the open position and vise versa.

Product
STOPCOCK EXT. SET W/ 3-WAY STOPCOCK WITH INJ. SITE AND SPINLOCK CONNECTOR, 22 inches and 37 inches
Manufacturer(s)
B. Braun Medical, Inc.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

B. Braun Medical, Inc. recalls STOPCOCK EXT. SET W/ 3-WAY STOPCOCK WITH INJ. SITE AND SPINLOCK CONNECTOR, 22 inches and 37 inches | RecallWatchUSA