Home / Recalls / Medical Devices
B-K Medical A/S recalls BK Medical ApS Ultrasound System Scanner bk2300 - Model #: bk3500 with software releases 1.2.0, 1.2.1 and 1.2.2. The s
Recalled August 10, 2016 · FDA recall Z-2325-2016 · B-K Medical A/S
Hazard
In the system (bk3500) software generic volume measurement (HxWxLxl) function the user may be unaware of the calculation factor (1) and assume the generic volume calculation factor is similar to that of the other (prostate, adenoma, testis, kidney, and bladder) volume calculation factors provided in the software. · FDA Class II
Units
6
Sold at
Worldwide Distribution - US to TX, FL, and MA. Internationally to Australia
Description
In the system (bk3500) software generic volume measurement (HxWxLxl) function the user may be unaware of the calculation factor (1) and assume the generic volume calculation factor is similar to that of the other (prostate, adenoma, testis, kidney, and bladder) volume calculation factors provided in the software.
Product
BK Medical ApS Ultrasound System Scanner bk2300 - Model #: bk3500 with software releases 1.2.0, 1.2.1 and 1.2.2. The system is intended for diagnostic ultrasound imaging or fluid flow analysis of the human body, data processing and guidance of puncture and biopsy. The system performs simple geometric measurements and calculations in the following areas: Emergency Medicine, Anesthesia, MSK, Vascular, Cardiology, OB/GYN
Manufacturer(s)
B-K Medical A/S
Made in
Denmark
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →