Home / Recalls / Medical Devices
Bard Access Systems recalls Titanium Single Lumen Low-Profile port, with Pre-Attached open-Ended Silicone 6.6Fr Catheter. Product Code 0602180 Us
Recalled January 29, 2014 · FDA recall Z-0815-2014 · Bard Access Systems
Hazard
Bard Access Systems is recalling various Low Profile Port with 6.6 Fr Single Lumen Broviac Catheter because a small number of Silicone catheters in the may be undersized to 5.3 Fr instead of 6.6 Fr as stated in the labeling for the device. · FDA Class II
Units
200
Sold at
Worldwide Distribution - US nationwide, Belgium, and Canada.
Description
Bard Access Systems is recalling various Low Profile Port with 6.6 Fr Single Lumen Broviac Catheter because a small number of Silicone catheters in the may be undersized to 5.3 Fr instead of 6.6 Fr as stated in the labeling for the device.
Product
Titanium Single Lumen Low-Profile port, with Pre-Attached open-Ended Silicone 6.6Fr Catheter. Product Code 0602180 Usage: The 6.6 Fr Open-Ended Single-Lumen venous catheters is packaged into various Bard Acess Systems finished goods as Attachable or Preconnected catheters in BardPort Implanted Port Kits These products are used for patient therapies requiring repeated access to the vascular system. They are indicated for infusion of medications, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. Subject ports are not indicated to Power Injection. Port access is performed by percutaneous needle insertion using a non-coring needle. The system consists of two primary components: an injection port with a self-sealing silicone septum and a radiopaque silicone catheter.
Manufacturer(s)
Bard Access Systems
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →