RecallWatchUSA

Home / Recalls / Medical Devices

Bard Peripheral Vascular Inc recalls Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G

Recalled January 17, 2024 · FDA recall Z-0643-2024 · Bard Peripheral Vascular

Hazard
The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures. · FDA Class II
Units
1,243,937
Sold at
Worldwide - US Nationwide distribution including in the states of PR, IL, MA, GA, KY, CA, TX, PA, IA, MN, NH, MD, FL, ID, IN, AL, VA, CO, WI, NY, NV, OK, OH, WA, MI, MS, AZ, HI, SD, NJ, NE, MT, NC, DC, CT, WY, UT, TN, LA, NM, SC, VT, MO, ME, AR, OR, KS, RI, AK, WV, DE, ND, PR, GU and the countries of CA, BE, AU, KR, SG, CN, TH, HK, VI, VN, AR, JP, CL, IN, TW, CO, MX, DO, BR, BD, MM, PK, NP, PA.

Description

The internal diameter of the coaxial cannula may be smaller or larger than the external diameter of the biopsy needle, and the length of the cannula may exceed the stated length on the label. As a result, the biopsy needle may not fit properly into the coaxial cannula, preventing access to the target tissue, which could lead to tissue injury, longer and repeat procedures.

Product
Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G x 16CM; 1610MSK/ Mission Kit 16G x 10CM; 1616MSK/ Mission Kit 16G x 16CM; 1810MSK/ Mission Kit 18G x 10CM; 1816MSK/ Mission Kit 18G x 16CM; 1820MSK/ Mission Kit 18G x 20CM; 1825MSK/ Mission Kit 18G x 25CM; 2010MSK/ Mission Kit 20G x 10CM; 2016MSK/ Mission Kit 20G x 16CM; 2020MSK/ Mission Kit 20G x 20CM
Manufacturer(s)
Bard Peripheral Vascular Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Bard Peripheral Vascular Inc recalls Bard Mission Disposable Core Biopsy Instrument Kit (REF/Name): 1410MSK/ Mission Kit 14G x 10CM; 1416MSK/ Mission Kit 14G | RecallWatchUSA