RecallWatchUSA

Home / Recalls / Medical Devices

Bard Peripheral Vascular Inc recalls BD Recanalization System, REF: BDRECANSYSTEM

Recalled January 17, 2024 · FDA recall Z-0676-2024 · Bard Peripheral Vascular

Hazard
Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute maximum catheter use time has been exceeded, which may increase the risk of the device overheating or breaking. · FDA Class II
Units
403
Sold at
US: CA, OH, PA, NY, TX, NJ, MS, IL, GA, AL, AZ, NC, NV, PR, IN, MA, WI, KS, AR, FL, MD, IA, NE, LA, HI, MO, DE, DC, VA, NM, CO, KY, ME, MN

Description

Recanalization systems may be in a manufacturing mode that allows the system to continuously run without the typical 30 second pauses, and 4-minute warning/5-minute restart required messages to indicate the ten-minute maximum catheter use time has been exceeded, which may increase the risk of the device overheating or breaking.

Product
BD Recanalization System, REF: BDRECANSYSTEM
Manufacturer(s)
Bard Peripheral Vascular Inc

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov