RecallWatchUSA

Home / Recalls / Medical Devices

Baxter Healthcare Corp. recalls Four Lead Arthroscopic Irrigation Set, for use with ARTHROMATIC Plastic Containers, Product Code 2C4031 -Approximate Len

Recalled July 17, 2013 · FDA recall Z-1684-2013 · Baxter Healthcare

Hazard
A firm's internal investigation led to the discovery of pinholes in the packaging of the products. The pinholes are not likely to be detected by the end user prior to use and could compromise the sterile barrier properties of the packaging. · FDA Class II
Units
249,576 units
Sold at
US Nationwide Distribution

Description

A firm's internal investigation led to the discovery of pinholes in the packaging of the products. The pinholes are not likely to be detected by the end user prior to use and could compromise the sterile barrier properties of the packaging.

Product
Four Lead Arthroscopic Irrigation Set, for use with ARTHROMATIC Plastic Containers, Product Code 2C4031 -Approximate Length 99" (2.5. m) Usage: Baxter irrigation sets are intended for the delivery of irrigating solutions from the fluid container to the irrigation site during bladder irrigation or endoscopic procedures including cytoscopy, transurethral resection and arthroscopic procedures. the sets are intended for use with Baxter irrigating solutions in Uromatic or Arthromatic plastic containers.
Manufacturer(s)
Baxter Healthcare Corp.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov