Baxter Healthcare Corporation recalls Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300

Recalled August 14, 2024 · FDA recall Z-2512-2024 · Baxter Healthcare

Hazard
Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport. · FDA Class II
Units
400 units
Sold at
Worldwide distribution - US Nationwide and the country of Canada.

Description

Baxter Healthcare Corporation is issuing an Urgent Medical Device Correction for the PST 500 U Precision Surgical Table due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport.

Product
Baxter PST 500 U, Precision Surgical Table, Part number (product code) 4080300
Manufacturer(s)
Baxter Healthcare Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov