Baxter Healthcare Corporation recalls Beach Chair Clamp, BEACH CHAIR SHLDER POSITIONER CMPT, REF A-90013, #800104. Component to position and support the pat

Recalled November 9, 2022 · FDA recall Z-0178-2023 · Baxter Healthcare

Hazard
the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity. · FDA Class II
Units
13 units
Sold at
MI, BA, WI, WA, TX, and Germany, Japan, South Korea, Sweden, South Africa

Description

the adhesive, applied to screws and bolts as an additional holding force, may not cure correctly, and as a result, could result in a loss of holding and reduced mechanical integrity.

Product
Beach Chair Clamp, BEACH CHAIR SHLDER POSITIONER CMPT, REF A-90013, #800104. Component to position and support the patients head in surgical procedures.
Manufacturer(s)
Baxter Healthcare Corporation

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov