Home / Recalls / Medical Devices
Baxter Healthcare Corporation recalls ONE-LINK Needle-free IV Connector, Product Codes: 7N8300, 7N8301, 7N8310, 7N8330K, 7N8332K, 7N8334K, 7N8370K, 7N8371, 7N
Recalled January 15, 2020 · FDA recall Z-0794-2020 · Baxter Healthcare
Hazard
Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector. · FDA Class II
Units
43,495,400 units
Sold at
Worldwide distribution - US Nationwide distribution and countries of Bermuda, Bahamas, Trinidad and Tobago, Grand Cayman, St. Thomas, Guyana and Barbados, Mexico, Canada, and New Zealand.
Description
Continu-Flo sets with male Luer bodies at the lower end of, and within, the ISO specification for outer diameter require a higher than expected connection force to fully inset and securely tighten the Luer to the ONE-LINK connector.
Product
ONE-LINK Needle-free IV Connector, Product Codes: 7N8300, 7N8301, 7N8310, 7N8330K, 7N8332K, 7N8334K, 7N8370K, 7N8371, 7N8375K, 7N8376K, 7N8377, 7N8378, 7N8390, 7N8391, 7N8399 - Product Usage: intended for single patient use with a vascular access device for the administration of drugs and solutions without needles, thus eliminating the potential for needle-stick injuries during use. This device is an in-line injection site which can be connected to standard male Luer adapters (e.g., syringes or sets) for the continuous or intermittent fluid administration or the withdrawal of fluids.
Manufacturer(s)
Baxter Healthcare Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →