Baxter Healthcare Corporation recalls Spectrum IQ Infusion Pump, Product Code 3570009

Recalled September 3, 2025 · FDA recall Z-2415-2025 · Baxter Healthcare

Hazard
Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy. · FDA Class II
Units
6 units
Sold at
US distribution to states of: AZ, FL

Description

Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

Product
Spectrum IQ Infusion Pump, Product Code 3570009
Manufacturer(s)
Baxter Healthcare Corporation