Baxter Healthcare Corporation recalls Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fl
Recalled August 18, 2021 · FDA recall Z-2110-2021 · Baxter Healthcare
Hazard
System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors. · FDA Class I
Units
130135 units
Sold at
United States (including Puerto Rico), Canada, Bahamas, Bermuda, Guyana, and Trinidad and Tobago.
Description
System errors were noted following changes the customers implemented to the configuration of their network and server systems. Gateway Server System performance anomalies resulted in excessive Spectrum IQ pump connectivity errors.
Product
Spectrum IQ Infusion System with Dose IQ Safety Software, Product Code 3570009. for the controlled administration of fluids.
Manufacturer(s)
Baxter Healthcare Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →