Home / Recalls / Medical Devices
Baxter Healthcare Corporation recalls Welch Allyn Life2000 Ventilation System, REF MS-01-0118, within exterior carton codes: 1) BT-20-0002, 2) BT-20-0002A, 3
Recalled October 30, 2024 · FDA recall Z-0046-2025 · Baxter Healthcare
Hazard
The Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy. · FDA Class I
Units
10,957 units
Sold at
US nationwide distribution.
Description
The Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy.
Product
Welch Allyn Life2000 Ventilation System, REF MS-01-0118, within exterior carton codes: 1) BT-20-0002, 2) BT-20-0002A, 3) BT-20-0002AP, 4) BT-20-0007, 5) BT-20-0011, and 6) RMS010118CP
Manufacturer(s)
Baxter Healthcare Corporation
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →