Bayer Healthcare, LLC recalls Contour Next Control Solution for Contour Next Blood Glucose Test System. Liquid quality control solution. Plastic bot
Recalled September 9, 2015 · FDA recall Z-2482-2015 · Bayer Healthcare
Hazard
If the control solution bottle is not thoroughly shaken prior to use, the control test result may be higher than the control range and may not be recognized and marked as a control test by the blood glucose meter. Also, the control solution, used as a quality control test in blood glucose meter systems, may read outside the specified range too high or too low. · FDA Class II
Units
5 million bottles of control solution
Sold at
Worldwide Distribution -- USA, Australia, Austria, Belgium, Canada, Croatia, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Portugal, Slovenia, Spain, Sweden, Switzerland, UK, Albania, Azerbaijan, Bulgaria, Czech Republic, Estonia, Hungary, Korea, Latvia, Lithuania, Philippines, Romania, Serbia, Slovakia, and Thailand.
Description
If the control solution bottle is not thoroughly shaken prior to use, the control test result may be higher than the control range and may not be recognized and marked as a control test by the blood glucose meter. Also, the control solution, used as a quality control test in blood glucose meter systems, may read outside the specified range too high or too low.
Product
Contour Next Control Solution for Contour Next Blood Glucose Test System. Liquid quality control solution. Plastic bottle, 2.5mL. Indicated for use with fresh fingertip capillary whole blood samples.
Manufacturer(s)
Bayer Healthcare, LLC
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →