Bayer Healthcare recalls Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuc
Recalled January 20, 2016 · FDA recall Z-0627-2016 · Bayer Healthcare
Hazard
The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail · FDA Class II
Units
198,598
Sold at
Nationwide Distribution to the states of : AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MI, MN, MO, NC, NE, NH, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, VA, VT and WI.
Description
The firm's investigation of the Medrad Intego indicates that the inner diameter of the saline tubing is out of specification and may result in the system failing to position the dose correctly during extraction, potentially leading to a low measurement of the extracted dose or to a higher dose being extracted. The occluded SAS tubing may impair the priming function and prompt an RP prime fail
Product
Medrad Source Administration Sets, Catalog number INT CSS. For use with MEDRAD Intego PET Infusion pump Systems, in nuclear medicine diagnostic procedures.
Manufacturer(s)
Bayer Healthcare
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →