BD Biosciences, Systems & Reagents recalls BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.
Recalled April 8, 2015 · FDA recall Z-1370-2015 · BD Biosciences, Systems & Reagents
Hazard
The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation. · FDA Class II
Units
72 devices
Sold at
Worldwide Distribution -US including TX, NJ, and OK and Internationally to Hong Kong, Japan, Singopore, and Belguim
Description
The circuit breakers on some BD FACSCalibur instruments may have the incorrect rated amperage (20 amps instead of 2 amps) causing the circuit breaker to fail to open in an over-current situation.
Product
BD FACSCalibur; Catalog Numbers 342973, 342975, 34976 In vitro Diagnostics Use.
Manufacturer(s)
BD Biosciences, Systems & Reagents
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →