Home / Recalls / Medical Devices
Becton, Dickinson and Company, BD Biosciences recalls BD FACSLyric Flow Cytometer, 3-Laser 10 Color Instrument, REF: 662878, and 66287808 when using BD Trucount Tubes
Recalled October 16, 2019 · FDA recall Z-0069-2020 · Becton, Dickinson and Company, BD Biosciences
Description
Flow cytometers, when used with specific tubes may experience excessive electronic abort counts (electronic abort count of beads exceeds 1.0% of the total processed events), which may affect the ratio of cell population events to bead events, potentially resulting in falsely high absolute counts. This may affect the following: initiation of prophylactic therapy for Opportunistic Infections (HIV patients), decisions regarding care/management (PID/immune suppressed patients), laboratory developed tests/experiments, specimen engraftment potency, and use of leucoreduced blood products for transfusion.
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →