RecallWatchUSA

Home / Recalls / Medical Devices

Becton, Dickinson and Company, BD Biosciences recalls BD" Mouse IgG2a Isotype Control APC X39 ASR

Recalled September 1, 2021 · FDA recall Z-2333-2021 · Becton, Dickinson and Company, BD Biosciences

Hazard
Complaints have been received for performance issues related to high background and staining. this issues is related to a manufacturing process error that can potentially cause a false positive staining due to contamination with another antibody. · FDA Class II
Units
23 reagents
Sold at
Worldwide distribution - U.S. Nationwide distribution in the states of CT, FL, KY, NY, and PA. The country of Taiwan.

Description

Complaints have been received for performance issues related to high background and staining. this issues is related to a manufacturing process error that can potentially cause a false positive staining due to contamination with another antibody.

Product
BD" Mouse IgG2a Isotype Control APC X39 ASR
Manufacturer(s)
Becton, Dickinson and Company, BD Biosciences

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Becton, Dickinson and Company, BD Biosciences recalls BD" Mouse IgG2a Isotype Control APC X39 ASR | RecallWatchUSA