Becton Dickinson & Co. recalls BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267
Recalled December 21, 2022 · FDA recall Z-0473-2023 · Becton Dickinson &
Hazard
Boxes contain a reduced concentration of Nicotinamide Adenine Dinucleotide and does not provide enough of the blood factors necessary to support Haemophilus growth. The reduction could cause a false negative result for Haemophilus isolation in a clinical specimen. · FDA Class II
Units
9,600 EA
Sold at
Worldwide distribution - US Nationwide and the countries of Canada, India, Philippines.
Description
Boxes contain a reduced concentration of Nicotinamide Adenine Dinucleotide and does not provide enough of the blood factors necessary to support Haemophilus growth. The reduction could cause a false negative result for Haemophilus isolation in a clinical specimen.
Product
BBL Chocolate II Agar- IVD for the isolation and cultivation of Neisseria species. Catalog (Ref) Number: 221267
Manufacturer(s)
Becton Dickinson & Co.