Becton Dickinson & Co. recalls BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitat

Recalled December 11, 2024 · FDA recall Z-0621-2025 · Becton Dickinson &

Hazard
BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility · FDA Class II
Units
13,190 units
Sold at
Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia.

Description

BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility

Product
BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263
Manufacturer(s)
Becton Dickinson & Co.