RecallWatchUSA

Home / Recalls / Medical Devices

Becton Dickinson & Co. recalls BD Synapsys Informatics Solution, Catalog Number 444150

Recalled July 12, 2023 · FDA recall Z-2099-2023 · Becton Dickinson &

Hazard
There is a risk of mis-association of patient demographic data for users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (Skip, Final Discard, Mark as Read, Save) to process patient results rapidly. · FDA Class II
Units
102 systems
Sold at
US States: CA, CO, IA, IL, MI, MN, MO, NY, OH, OK, SC, TN, TX, UT, WI. Australia, Austria, Belgium, Canada, Denmark, France, Germany, Italy, Japan, Luxembourg, Netherlands, New Zealand, Norway, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, United Kingdom

Description

There is a risk of mis-association of patient demographic data for users who choose to report results manually when their Service Control Unit (SCU) is operating near capacity and are using save actions (Skip, Final Discard, Mark as Read, Save) to process patient results rapidly.

Product
BD Synapsys Informatics Solution, Catalog Number 444150
Manufacturer(s)
Becton Dickinson & Co.

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov