RecallWatchUSA

Home / Recalls / Medical Devices

Becton Dickinson & Company recalls BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (4.5 ml x 13 x 75 mm) Catalog Number: 369714 Pro

Recalled February 22, 2017 · FDA recall Z-1189-2017 · Becton Dickinson &

Hazard
A portion of this lot of blood collection tubes was manufactured with approximately twice the expected amount of buffered sodium citrate additive, an anticoagulant. · FDA Class II
Units
1,488,800 tubes
Sold at
Worldwide Distribution - US Nationwide and the countries of Canada, Mexico, Indonesia

Description

A portion of this lot of blood collection tubes was manufactured with approximately twice the expected amount of buffered sodium citrate additive, an anticoagulant.

Product
BD Vacutainer Buffered Sodium Citrate: (9NC) Blood Collection Tube (4.5 ml x 13 x 75 mm) Catalog Number: 369714 Product Usage: The BD Vacutainer Safety Coagulation tube is an evacuated blood collection tube that provides means of collecting, transporting and processing blood in a closed tube. The buffered sodium citrate additive provides an anti-coagulated specimen that may be used for clinical laboratory coagulation assays and determinations
Manufacturer(s)
Becton Dickinson & Company

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov