Becton Dickinson & Company recalls BD Vacutainer C&S Transfer Straw Kit C&S Preservative Plus Urine Tube. 4.0 ml, 13 X 75 mm STERILE REF 364953. Becton

Recalled April 10, 2013 · FDA recall Z-1056-2013 · Becton Dickinson &

Hazard
The BD Vacutainer Urinalysis Transfer Straw Kit #364991, Lot 1335484 has preprinted polybags which are incorrectly labeled as Catalog #364953. BD advises any Customer who purchased this Kit that the Tube in this Kit is not a C&S tube. If a Customer relies solely on the polybag labeling and uses the Tube inside the Kit for the purpose of collecting urine for culture and sensitivity, it is possible · FDA Class III
Units
236,400 units
Sold at
Nationwide Distribution including Arizona, California, Colorado, Georgia, Florida, Illinois, Iowa, Kentucky, Louisiana, Maryland, Massachussets, Michigan, Minnesota, Missouri, New Jersey, New Mexico, New York, North Carolina, Ohio, Oregon, Pennsylvania, Tennessee, Texas, Utah, Virginia, Washington DC, Washington, West Virginia, and Wisconsin.

Description

The BD Vacutainer Urinalysis Transfer Straw Kit #364991, Lot 1335484 has preprinted polybags which are incorrectly labeled as Catalog #364953. BD advises any Customer who purchased this Kit that the Tube in this Kit is not a C&S tube. If a Customer relies solely on the polybag labeling and uses the Tube inside the Kit for the purpose of collecting urine for culture and sensitivity, it is possible

Product
BD Vacutainer C&S Transfer Straw Kit C&S Preservative Plus Urine Tube. 4.0 ml, 13 X 75 mm STERILE REF 364953. Becton Dickinson & Company, Franklin Lakes, NJ USA, The kit consists of a transfer straw and urinalysis tube intended for UA chemistry urine testing.
Manufacturer(s)
Becton Dickinson & Company

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Becton Dickinson & Company recalls BD Vacutainer C&S Transfer Straw Kit C&S Preservative Plus Urine Tube. 4.0 ml, 13 X 75 mm STERILE REF 364953. Becton | RecallWatchUSA