Becton Dickinson & Company recalls BD Vacutainer¿ EDTA Tan Top Tube; Catalog Number: 367855 (UDI: 50382903678551)
Recalled March 28, 2018 · FDA recall Z-1140-2018 · Becton Dickinson &
Hazard
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology. · FDA Class I
Units
3,956,900
Sold at
Distributed in 46 states: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NY, NV, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Puerto Rico.
Description
BD Vacutainer EDTA Lavender, Tan, and Pink Top Tubes and BD Vacutainer Lithium Heparin Green Top Tubes can cause an underestimation of lead in blood samples when used with Magellan Diagnostics LeadCare assays, employing the Anodic Stripping Voltammetry (ASV) methodology, or any other assay employing ASV methodology.
Product
BD Vacutainer¿ EDTA Tan Top Tube; Catalog Number: 367855 (UDI: 50382903678551)
Manufacturer(s)
Becton Dickinson & Company
Source: U.S. Food and Drug Administration. Official recall notice on fda.gov →