RecallWatchUSA

Home / Recalls / Medical Devices

Belmont Instrument LLC recalls Allon 2001. Part Number: 200-00263

Recalled June 18, 2025 · FDA recall Z-1966-2025 · Belmont Instrument

Hazard
An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition. · FDA Class II
Units
229 units
Sold at
Worldwide - US Nationwide distribution in the states of CA, DC, LA, MA, NM, PA, TX, VA, WA and the countries of Belgium, Bolivia (Plurinational State of), Brazil, Canada, Colombia, Israel, Italy, Portugal, Romania, Sweden, Taiwan (Province of China), Thailand, United Kingdom.

Description

An investigation into a report of the patient wrap (ThermoWrap), used with the Allon device, exceeding the set temperature determined that if: 1. The Human Sensor Board is disconnected from the Controller Board in the Allon device AND the temperature alarms are disabled, or 2. The Water In and Water Out and Thermostat sensors are disconnected from the Human Sensor Board AND the temperature alarms are disabled, the device will proceed to heat the water above the hardware fault limit without notifying the user of the faulty condition.

Product
Allon 2001. Part Number: 200-00263
Manufacturer(s)
Belmont Instrument LLC

Source: U.S. Food and Drug Administration. Official recall notice on fda.gov

Belmont Instrument LLC recalls Allon 2001. Part Number: 200-00263 | RecallWatchUSA